• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PKR FEMUR #6 LM/RL; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PKR FEMUR #6 LM/RL; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5610-F-601
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Ambulation or Postural Difficulties (2544)
Date of Event 12/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned to the manufacturer.
 
Event or Problem Description
The patient received a removal of the sled prosthesis due to aseptic loosening on (b)(6) 2021.During removal of the sled prosthesis, it was noticed that the femoral part was broken.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a triathlon femoral component was reported.The event was confirmed via evaluation of the returned device and through clinician review of provided medical records.Method & results: -product evaluation and results: visual inspection: the pka femoral component was returned.There is a layer of bone cement on the underside of the device.The device has fractured between the 2 pegs.The damage is consistent with fatigue accumulation ultimately ending in a fatigue fracture of the device requiring revision.-clinician review: a review of the provided medical information by a clinical consultant indicated: "this inquiry concerns a patient who underwent a medial compartment partial knee arthroplasty in 2010 and subsequently required revision surgery in 2021.Initially it was felt that revision was needed for tibial loosening.At the time of the revision, it was found that the femoral component was fractured.In my opinion, the femoral fracture could have been seen prior to the revision.It was present on the lateral x-ray provided.I can confirm that this event happened since i was able to see the x-ray and intraoperative photo showing the fractured femoral component.The loosening of the tibial component is less clear shown.The root cause of this event cannot be determined with certainty.Causes of aseptic loosening after 10 years are multifactorial.These include surgical technique factors and patient factors, including bmi and activity level.Causes of femoral component fracture are also multifactorial.Fracture can occur by fatigue if fixation was solid in one area and not in another.Surgical technique therefore can be contributory.In this particular case i saw very little cement if any anteriorly and no cement posteriorly on the femoral component.This possibly could cause a fatigue effect resulting in fracture." -product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: it was reported that the patient was revised due to tibial loosening, and, intraoperatively, fracture of the femoral component was observed.The pka femoral component was returned for evaluation.There is a layer of bone cement on the underside of the device.The device has fractured between the 2 pegs.The damage is consistent with fatigue accumulation ultimately ending in a fatigue fracture of the device requiring revision.A review of the provided medical information by a clinical consultant indicated "the femoral fracture could have been seen prior to the revision.It was present on the lateral x-ray provided.I can confirm that this event happened since i was able to see the x-ray and intraoperative photo showing the fractured femoral component.Causes of femoral component fracture are also multifactorial.Fracture can occur by fatigue if fixation was solid in one area and not in another.Surgical technique therefore can be contributory.In this particular case i saw very little cement if any anteriorly and no cement posteriorly on the femoral component.This possibly could cause a fatigue effect resulting in fracture." no further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
The patient received a removal of the sled prosthesis due to aseptic loosening on (b)(6) 2021.During removal of the sled prosthesis, it was noticed that the femoral part was broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PKR FEMUR #6 LM/RL
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13569308
Report Number0002249697-2022-00253
Device Sequence Number3860655
Product Code MBH
UDI-Device Identifier07613327027822
UDI-Public07613327027822
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K071881
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number5610-F-601
Device Catalogue Number5610-F-601
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2022
Supplement Date Received by Manufacturer08/25/2022
Initial Report FDA Received Date02/21/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
Patient SexUnknown
Patient Weight106 KG
-
-