DEPUY INTERNATIONAL LTD - 8010379 9/10 COCR HEAD 22.225MM -3; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1521-90-051
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2021, the patient underwent the tha surgery with the stem and head implant in question.A dual mobility produced by zimmer biomet was used for cup.After the surgery, the patient had frequent dislocations on an unknown date.On (b)(6) 2022, the revision surgery was performed.Stem was replaced with new one (16×11×30+4neck).The sleeve was not removed and continued to be used.Head implant was replaced with new same one.No further information is available.Doi: (b)(6) 2021.Dor: (b)(6) 2022.Unknown hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9/10 COCR HEAD 22.225MM -3
Common Device Name
SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13569557
Report Number1818910-2022-03211
Device Sequence Number5950701
Product Code JDI
UDI-Device Identifier10603295276562
UDI-Public10603295276562
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K871867
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1521-90-051
Device Catalogue Number152190051
Device Lot NumberD19091533
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2022
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date02/21/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
S-ROM*SLEEVE PRX ZTT, 16B-LRG.; SROM 9/10 16X11X130 30.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-