| Model Number |
1012452-12 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported the procedure was to treat the right circumflex.A 3.75x12 mm nc trek balloon dilatation catheter was used but did not deflate properly.They were able to eventually get it deflated but it took awhile and didn't behave appropriately.There were no adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported deflation issue.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|