| Model Number |
G5A-80-L |
| Medical Device Problem Codes |
High impedance (1291); Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device displayed a "check pads" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device did not detect the attached electrode pads and failed for high impedance.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section b5 and h6 (device problem code).Evaluation: the customer's report was observed during review of the device data logs.The log was cleared prior to evaluation.The device was put through extensive testing using test pads without duplicating the report.An internal inspection found no discrepancies.The device was recertified and returned to the customer.The electrode pads used at the time of the report were not returned as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
|
| |
|