Model Number 24674
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2021
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on 04feb2022.It was reported that inflation issue occurred.The target lesion was located in the iliac vein.A 12.0 x 80, 135cm mustang balloon catheter was advanced for dilatation.However, during the procedure, the balloon could not be deployed properly after pressure.The device was removed from the patient's body and the procedure was completed with another of same device.No patient complications were reported and the patient status was stable.However, device analysis revealed balloon pinhole.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: mustang 12.0 x 80, 135cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.Blood was identified within the balloon which is evidence of a device leak.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 13mm proximal of the proximal markerband.A visual and microscopic examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no issues or damage to the tip of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13573127
Report Number2134265-2022-01384
Device Sequence Number5912553
Product Code FGE
UDI-Device Identifier08714729795285
UDI-Public08714729795285
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/11/2024
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0026609536
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2022
Initial Report FDA Received Date02/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age38 YR
Patient SexMale
Patient Weight75 KG
-
-