| Model Number |
SE5425MVB |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Vitreous Hemorrhage (2143)
|
| Date of Event |
01/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was discarded after surgery, and is therefore not available for evaluation.Manufacturing and sterilization records for this lot were reviewed and found to be acceptable.The self-sealing ability of a sclerotomy wound is affected by wound architecture, scleral tunnel length, scleral elasticity, wound apposition by residual vitreous, surface tension of a gas bubble, and intraocular pressure.Leakage from sclerotomies is more likely in highly myopic eyes with low scleral rigidity, in eyes with scarred conjunctiva or sclera from previous surgery, in marfan''s syndrome, and in young children.Partial fluid-air exchange may help reduce wound leak from the sclerotomies until fibrin seals the wounds.Procedure causality is "probable".The surgeon will suture 25-gauge sclerotomy sites henceforth.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
A user facility reports performing a vitrectomy using stellaris elite 25g posterior pack due to a post-op vitreous hemorrhage.Right after surgery, the patient was stable without any leakage/ bleeding, normal iop pre-op and right after surgery.No sutures were used.There were no complications during surgery, no post-op inflammation/ infection.No post-op trauma to the eye.The next day in clinic the surgeon noticed a problem with the intraocular pressure.The surgeon found the patient had post-op, sub-conjunctival bleeding around the infero-temporal site, and active leaking at the sup-temporal sclerotomy area where fluid is leaking through the superior-temporal conjunctival incision.The iop was 8).Air-fluid exchange was performed.The leaks stopped after partial air-fluid exchange.The patient outcome is normal intraocular pressure.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing and sterilization records were reviewed and found to be acceptable.The product root cause could not be determined as the sample was not returned and evaluated.The trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.There is currently a capa opened on this product for leaking valves.The investigation is complete.
|
| |
|
Search Alerts/Recalls
|