| Model Number |
24672 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: a mustang 6.0 x 120, 135cm was returned for analysis.A visual examination identified that the balloon was subjected to positive pressure.The returned device was attached to a boston scientific encore inflation unit.Positive pressure was applied using glycerol/water and a pinhole leak was identified 4mm distal from the proximal markerband.A visual examination found no issue with the tip section of the device.A visual examination found no issue with the markerbands.A visual and tactile examination identified no kinks or damage to the shaft of the device.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 28-jan-2022.It was reported that balloon leak occurred.A 6.0 x 120, 135cm mustang balloon catheter was advanced for dilatation.However, during the procedure balloon would not inflate.After withdrawal it was found that there was a leakage of serum and contrast.The procedure was completed with another of the same device.No patient complications were reported.However, returned device analysis revealed a balloon pinhole.
|
| |
|