Model Number 1012274-30
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record and complaint history could not be conducted because the lot number was not provided.The investigation determined the reported difficulty removing the device, separations, additional treatment, and foreign body in patient appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.User facility medwatch form mw5106318.
 
Event or Problem Description
User facility medwatch report (mw5106318) received stating: when the device was being removed from the patient, it became stuck to an area of calcium in the patient's vessel.Ultimately dislodging the balloon shaft becoming retained.Additionally, it was reported that the 3.00 x 30mm trek balloon dilatation catheter (bdc) was advanced without resistance to the moderately calcified lesion in the profunda artery.The balloon was inflated twice and deflated without issue.During removal, resistance was noted due to calcification in the vessel and the distal shaft, including the balloon separated.Snaring was performed in an attempt to remove the separated portion of the device, but it was not successful, so a bare metal stent was advanced and implanted to secure the separated portion.There was no adverse patient sequela or a clinically significant delay in procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13573539
Report Number2024168-2022-01828
Device Sequence Number5680934
Product Code LOX
UDI-Device Identifier08717648138461
UDI-Public08717648138461
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1012274-30
Device Catalogue Number1012274-30
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2022
Initial Report FDA Received Date02/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-