Model Number 301803MS
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an unknown patient underwent an atrial fibrillation (afib) ablation procedure with a preface® guiding sheath with multipurpose curve.The hemostatic valve was leaking.It was reported that after inserting the brockenbrough (b.B) transseptal needle, there was a trace of reversed blood from the sheath.This occurred 40 minutes after the patient was entered.The issue was resolved by changing out the product.The hemostatic valve was intact, and the reversed blood volume was a few drops.There was no occurrence of health hazard due to reversed blood.No medical treatment was performed in association with the occurrence of reversed blood.The patient had no air contamination.The procedure was completed without patient's consequence.The patient¿s hemodynamics were not compromised due to bleeding.A few drops of blood were lost, but the exact amount is unknown.It was also reported that the reversed blood was coming from the hemostatic valve portion of the sheath.The reversed blood reported was not related with the bb.Hemostatic valve leak is mdr-reportable.
 
Additional Manufacturer Narrative
Initial reporter phone: (b)(6).The device has been reported as discarded, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.A device history record was reviewed for the finished device number 17982271, and no internal action related to the complaint was found during the review.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREFACE® GUIDING SHEATH WITH MULTIPURPOSE CURVE
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
gabriel alfageme
31 technology dr
irvine, CA 92618
9497898687
MDR Report Key13575200
Report Number2029046-2022-00349
Device Sequence Number14544990
Product Code DYB
UDI-Device Identifier10846835000030
UDI-Public10846835000030
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K982740
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model Number301803MS
Device Catalogue Number301803MS
Device Lot Number17982271
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2022
Initial Report FDA Received Date02/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BROCKENBROUGH TRANSSEPTAL NEEDLE
Patient SexUnknown
-
-