| Model Number |
R5100052 |
| Medical Device Problem Codes |
Device Alarm System (1012); Pumping Problem (3016)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: the reported issues of the instrument would not prime, and it displayed a possible error message of "pump stopped" were verified during field service.During service the technician noted that when turning on the instrument the motor would spin during the self-test.The technician pressed the start button for the pm4 which did not work.The technician then swapped the connectors on the sin-ui board, and they were able to turn the motor on with a pm2's start button while pm2 itself would not start with pm4's start button.The issues were resolved by replacing the overlay for pm4.Post-repair testing was performed per specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that prior to use of this performer instrument, the customer noticed that the instrument would not prime and it displayed a possible error message of "pump stopped".The instrument was changed out with a backup and there was no reported adverse patient effect.Additional information received confirms the hand crank was not required to maintain flow.
|
| |
|