• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CROSSER; RECANALIZATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CROSSER; RECANALIZATION CATHETER Back to Search Results
Model Number CRUOS106
Medical Device Problem Codes Difficult to Remove (1528); Retraction Problem (1536); Material Separation (1562); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that during a recanalization procedure in the right iliac with total occlusion, the device allegedly had compatibility issue.It was further reported that the device had withdrawal issue.The procedure was completed using another device.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a recanalization procedure in the right iliac, the device allegedly had compatibility issue.It was further reported that the device had retraction issue.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one 6s crosser cto recanalization catheter loaded onto a sidekick was returned for evaluation.Crosser can be removed from the sidekick catheter without issues.Based on the findings, the investigation is unconfirmed for the reported device-device incompatibility and retraction problem.A definitive root cause for the reported device-device incompatibility and retraction problem could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: g3, h6 (method) h11: g1, h6 (device, result, conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER
Common Device Name
RECANALIZATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13576904
Report Number2020394-2022-00121
Device Sequence Number5680545
Product Code PDU
UDI-Device Identifier10801741125611
UDI-Public(01)10801741125611
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K161208
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCRUOS106
Device Catalogue NumberCRUOS106
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/26/2022
Supplement Date Received by Manufacturer05/26/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date05/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
Patient Weight99 KG
-
-