H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one 6s crosser cto recanalization catheter loaded onto a sidekick was returned for evaluation.Crosser can be removed from the sidekick catheter without issues.Based on the findings, the investigation is unconfirmed for the reported device-device incompatibility and retraction problem.A definitive root cause for the reported device-device incompatibility and retraction problem could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: g3, h6 (method) h11: g1, h6 (device, result, conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|