• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 PRISTINE; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 PRISTINE; DIALYSIS CATHETER Back to Search Results
Model Number 5403230
Medical Device Problem Codes Component Missing (2306); Device Markings/Labelling Problem (2911); Component Misassembled (4004); Inaccurate Information (4051)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation as well as photos were provided for review.The investigation of the reported event is currently underway.(expiry date: 02/2023).
 
Event or Problem Description
It was reported that prior to a dialysis catheter placement procedure, the labeling of the catheter allegedly found to be mislabeled.The procedure was completed using another device.There was no patient contact.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation as well as photos were provided for review.The investigation of the reported event is currently underway.(expiry date: 02/2023).
 
Event or Problem Description
It was reported that prior to a dialysis catheter placement procedure, the labeling of the catheter allegedly found to be mislabeled.The procedure was completed using another device.There was no patient contact.
 
Event or Problem Description
It was reported that prior to a dialysis catheter placement procedure, the incorrect component was allegedly found in the package.The procedure was completed using another device.There was no patient contact.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one unsealed pristine 23cm catheter kit was returned for evaluation and two electronic photos were provided for review.Gross visual and dimensional evaluations were performed on the returned device.The dimensions measured during sample evaluation does not match the dimensions of the device (rm1923102) with the reported lot which is a 23cm catheter.Moreover the dimensions of the returned device matches with the material (rm1923104) which is a 33cm pristine catheter.However upon further review on the manufacturing, packaging, biometrics, printing and stocking process of the device with the particular lot shows no discrepancies noted related to incorrect sized device within package or incorrect specification printed on device.Therefore, the investigation is inconclusive for the reported inaccurate information, component misassembled and component missing issues as the device was returned with the packaging in unsealed condition and moreover the exact circumstances upon opening the package are unknown.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 02/2023), g3, h6 (method).H11: h6 (device, result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISTINE
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13576927
Report Number3006260740-2022-00421
Device Sequence Number12144136
Product Code MSD
UDI-Device Identifier00801741219047
UDI-Public(01)00801741219047
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K203767
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5403230
Device Catalogue Number5403230
Device Lot NumberREFT0662
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer03/17/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date03/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-