• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC COOK® MULTI-USE HOLMIUM LASER FIBER; GEX LASER INSTRUMENT, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK® MULTI-USE HOLMIUM LASER FIBER; GEX LASER INSTRUMENT, SURGICAL Back to Search Results
Model Number G25298
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported, a cook® multi-use holmium laser fiber did not work upon its second use.The procedure was not able to be completed.No unintended section of the device remained inside the patients body.No adverse effects have been reported due to this occurrence.Upon return and initial evaluation of the complaint device on (b)(6) 2022, the tip was fractured.
 
Additional Manufacturer Narrative
Customer (person) - phone: (b)(6).Device evaluated by mfg = other (81) - device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, a cook® multi-use holmium laser fiber did not work upon its second use.The procedure was not able to be completed.No unintended section of the device remained inside the patients body.No adverse effects have been reported due to this occurrence.Upon return and initial evaluation of the complaint device on 17feb2022, the tip was fractured.Investigation evaluation: reviews of the complaint history, device history record, manufacturing instructions, instructions for use (ifu) and specifications were conducted during the investigation.A visual inspection of the complaint device was also conducted.Examination of the returned complaint device revealed the laser fiber tip was broken with a jagged tip.There was no charring or discoloration on the blue jacket noted.A document-based investigation evaluation was performed.No related non-conformances were recorded, and there have been no other reported complaints for this lot number.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The instructions for use (ifu) provides the following information to the user related to the failure mode: "upon removal from package, inspect the product to ensure no damage has occurred." as the customer did not report any issue regarding the laser fiber tip, it is possible the damage to the laser fiber occurred as a result of returning it to cook for investigation.However, based upon the available information, a definitive cause could not be determined.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK® MULTI-USE HOLMIUM LASER FIBER
Common Device Name
GEX LASER INSTRUMENT, SURGICAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key13577455
Report Number1820334-2022-00276
Device Sequence Number14634106
Product Code GEX
UDI-Device Identifier10827002252988
UDI-Public(01)10827002252988(17)240609(10)14015906
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K133788
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/09/2024
Device Model NumberG25298
Device Catalogue NumberHLF-M273-HSMA
Device Lot Number14015906
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer03/15/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient Age60 YR
Patient SexUnknown
-
-