Model Number TCN-5
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during sterilization, blood was noted inside and was oozing out from the hub.
 
Event or Problem Description
It was reported that during sterilization, blood was noted inside and was oozing out from the hub.
 
Additional Manufacturer Narrative
Analysis of the returned tcn-5 lot 26115925 revealed that the electrode was found to have fluid ingress when the hub was soaked in water.Overuse of the electrodes (sterilizing units beyond the number of cycles the units are validated for) can lead to additional wear on the units and the epoxy inside the hub, leading to cracks that allow for fluid ingress.
 
Event or Problem Description
It was reported that during sterilization, blood was noted inside and was oozing out from the hub.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
PROBE, RADIOFREQUENCY LESION
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key13577616
Report Number3006630150-2022-00625
Device Sequence Number4368463
Product Code GXI
UDI-Device Identifier00813250013047
UDI-Public00813250013047
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K050084
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTCN-5
Device Catalogue NumberTCN-5
Device Lot Number26115925
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/28/2022
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date04/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/02/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;