• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW AGBA PICC/DELTA FG: 2-L 5.5 FR; CATHETER,INTRAVASCULAR,THERAPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW AGBA PICC/DELTA FG: 2-L 5.5 FR; CATHETER,INTRAVASCULAR,THERAPE Back to Search Results
Model Number IPN920209
Medical Device Problem Codes Difficult to Advance (2920); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
Customer reported picc line would not advance into the peel-away sheath.The distal end of the catheter advanced but they were unable to advance the entire picc.They were finally able to "force" the catheter through the sheath and complete the procedure.No patient harm reported.The patient's condition is reported as fine.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Customer reported picc line would not advance into the peel-away sheath.The distal end of the catheter advanced but they were unable to advance the entire picc.They were finally able to "force" the catheter through the sheath and complete the procedure.No patient harm reported.The patient's condition is reported as fine.
 
Additional Manufacturer Narrative
Qn# (b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed with one relevant finding identified.For material cdc-45552-cat2 and lot 13c21e1118, a non-conformance was initiated in regards to a stylet tight in picc catheter.The instructions for use (ifu) provided with this kit warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested.Do not use excessive force when introducing guidewire, peel-away sheath over tissue dilator, or tissue dilator as this can lead to venospasm, vessel perforation, bleeding, or component damage." without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
Customer reported picc line would not advance into the peel-away sheath.The distal end of the catheter advanced but they were unable to advance the entire picc.They were finally able to "force" the catheter through the sheath and complete the procedure.No patient harm reported.The patient's condition is reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW AGBA PICC/DELTA FG: 2-L 5.5 FR
Common Device Name
CATHETER,INTRAVASCULAR,THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key13577849
Report Number9680794-2022-00111
Device Sequence Number5644332
Product Code LJS
UDI-Device Identifier10801902192568
UDI-Public10801902192568
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K153487
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/31/2023
Device Model NumberIPN920209
Device Catalogue NumberASK-45552-CRMC
Device Lot Number13F21L0426
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2022
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-