STRYKER SPINE-US REFLEX HYBRID SCREW EXTRACTOR INNER SHAFT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 48511905R
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a reflex-hybrid fixation pin and screw extractor inner shaft were observed to be bent outside the or.There is no procedure associated with this reported event, no patient involvement, and no adverse consequence has been reported.This report will capture the inner shaft.
 
Event or Problem Description
It was reported that a reflex-hybrid fixation pin and screw extractor inner shaft were observed to be bent outside the or.There is no procedure associated with this reported event, no patient involvement, and no adverse consequence has been reported.This report will capture the inner shaft.
 
Additional Manufacturer Narrative
The event was confirmed via visual inspection of the returned device.The body of the inner shaft is deformed / bent.Threading at the tip of the instrument is severely deformed / worn.The device is no longer functional.Manufacturing history was reviewed and no issues were identified.The lot was manufactured in jun of 2014.A review of complaint history associated with the subject catalog number was performed, and no adverse trends were observed.No additional information was provided from a complainant.The general instrument ifu states that, " the life of the instrument depends on the number of times they are used as well as the precautions taken in handling, cleaning and storage.Great care must be taken of the instruments to ensure that they remain in good working order." the cause of the reported event is normal wear due to device being is use for eight years.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFLEX HYBRID SCREW EXTRACTOR INNER SHAFT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
Manufacturer (Section G)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR   33610
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key13577967
Report Number0009617544-2022-00020
Device Sequence Number4555649
Product Code LXH
UDI-Device Identifier07613327276831
UDI-Public07613327276831
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48511905R
Device Catalogue Number48511905R
Device Lot Number130938
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer05/11/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/24/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1