• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RXSIGHT, INC. LIGHT ADJUSTABLE LENS (LAL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RXSIGHT, INC. LIGHT ADJUSTABLE LENS (LAL) Back to Search Results
Model Number 60005
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Visual Impairment (2138)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The site reported that the light adjustable lens was explanted from a patient.No additional information was provided as to the reason for the explant.Rxsight's first awareness was (b)(6) 2022.The device history record was reviewed.No issues were noted.The explanted lens was returned for evaluation.The lens was cut into multiple fragments.Visual and optical testing of the lens found no issues with the lens.
 
Event or Problem Description
The site reported that the light adjustable lens was explanted from a patient.No additional information was provided as to the reason for the explant.Rxsight's first awareness was (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT ADJUSTABLE LENS (LAL)
Common Device Name
LIGHT ADJUSTABLE LENS (LAL)
Manufacturer (Section D)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer (Section G)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer Contact
kelli carrillo
100 columbia
aliso viejo, CA 92656
MDR Report Key13577992
Report Number3012712027-2022-00006
Device Sequence Number4673657
Product Code PZK
UDI-Device Identifier00818806020043
UDI-Public01008188060200431120112117231031
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P160055
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model Number60005
Device Catalogue Number60005
Device Lot NumberL02-001938
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/25/2022
Initial Report FDA Received Date02/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-