| Model Number |
89-10536 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Pack contained suction tubing that appeared to have a brownish substance on what looks like the outside of the tubing.The patient was not in the room.The offending item along with the entire pack was removed.A new pack was brought in and carefully inspected prior to bringing the patient in the room.
|
| |
|
Additional Manufacturer Narrative
|
|
A user facility medwatch report (# (b)(4)) was received on 02/14/2022 reporting pack contained suction tubing that appeared to have a brownish substance on what looks like the outside of the tubing.The patient was not in the room.A supplier corrective action request (scar) was issued to the tubing supplier (b)(4).The sample has not been returned to deroyal for evaluation at this time.This investigation is ongoing at this time.No further information is available at this time.We will provide follow-up report if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
The tubing was returned to deroyal for evaluation.A supplier corrective action request (scar) was issued to the tubing supplier conmed.Potential root cause: raw material (resin) with foreign matter from conmed's supplier.Cleaning routing not determined in the header, mesh and the die.The following corrective actions have taken place by conmed: requested an incoming inspection of the resin.Updated orders to include the cleaning routing process to include the period of time of the change frequency, this includes changes in the header, the meshes, the dies as applied.The burned resin will be removed before it accumulates in the die.Deroyal no longer receives this product from conmed.This investigation is complete at this time.No further information is available at this time.We will provide follow-up report if additional information becomes available.
|
| |
|
Search Alerts/Recalls
|