The suspect device has been returned to olympus for evaluation and the investigation is in process.The olympus service center confirmed the failed probe check error code u509.Both switches were checked and found both switches are functional.A visual inspection on the received condition was performed on the device; there is no tissue build up.The teflon pad was inspected and found it is in good condition.The distal end of the device was inspected under a microscope and found approximately 12mm of the probe is detached; detached probe was not returned.The wiper movement is normal.The consistency of movement of the handle and jaw is normal.The handle load is normal.The rotation of the knob torque is normal and smooth.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the probe broke due to contact with a surgical instrument or the non-insulated area of grasping section.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1.When seal & cut output at pre-use inspection, the probe came in contact with a metal or a surgical instrument.Consequently, a scratch was made on the probe.2.The probe received an output load in seal & cut mode.As a result, the probe cracked from the scratch and the error occurred.3.The probe broke when added load in pre-use inspection.This following information is included in the instructions for use (ifu): "do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities." olympus will continue to monitor the performance of this device.
|