• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0545FCS
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device has been returned to olympus for evaluation and the investigation is in process.The olympus service center confirmed the failed probe check error code u509.Both switches were checked and found both switches are functional.A visual inspection on the received condition was performed on the device; there is no tissue build up.The teflon pad was inspected and found it is in good condition.The distal end of the device was inspected under a microscope and found approximately 12mm of the probe is detached; detached probe was not returned.The wiper movement is normal.The consistency of movement of the handle and jaw is normal.The handle load is normal.The rotation of the knob torque is normal and smooth.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event or Problem Description
A user facility reported to olympus the thunderbeat produced an unspecified error message while doing the probe check during preparation for use of the device.Upon inspection and testing of the customer returned device, it was observed the probe is detached and not returned with the unit.This report is being submitted for the malfunction found during evaluation.There was no harm or user injury reported due to the event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the probe broke due to contact with a surgical instrument or the non-insulated area of grasping section.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1.When seal & cut output at pre-use inspection, the probe came in contact with a metal or a surgical instrument.Consequently, a scratch was made on the probe.2.The probe received an output load in seal & cut mode.As a result, the probe cracked from the scratch and the error occurred.3.The probe broke when added load in pre-use inspection.This following information is included in the instructions for use (ifu): "do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities." olympus will continue to monitor the performance of this device.
 
Additional Manufacturer Narrative
Correction to g3 of the initial medwatch.The aware date should be 11-nov-2021.Investigation activities have been opened to manage the actions related to this report and any required mdr reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP TYPE S
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13578144
Report Number8010047-2022-03318
Device Sequence Number14863933
Product Code GEI
UDI-Device Identifier04953170383519
UDI-Public04953170383519
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K211838
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTB-0545FCS
Device Lot NumberKR110725
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2022
Supplement Date Received by Manufacturer03/16/2022
11/08/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date04/13/2022
12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-