MAUDE Adverse Event Report: EXACTECH, INC. ACUMATCH GXL 15DEG LINER 32MM SZ H; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. ACUMATCH GXL 15DEG LINER 32MM SZ H; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number ACUMATCH GXL 15DEG LINER 32MM SZ H
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 01/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately sixteen years post initial tha, this male patient had a revision due to osteolysis.There is evidence of poly wear.Patient current status is stable.The devices are not available for evaluation.
 
Additional Manufacturer Narrative
Concomitant medical products: p-series pf plasma collarless sz 4 12/14 (cat# 118-01-04 / serial# (b)(4)), acumatch cluster cup porous coated 56mm (cat# 120-01-56 / serial#(b)(4)), acumatch gxl 15deg liner 32mm sz h (cat# 132-32-28 / serial# (b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections d10, g3, g6, h2, h3 and h6 have been updated accordingly.Section d10: concomitant medical product, 12/14 zirconia head 32mm +3.5mm neck (cat# 148-32-03 / serial# (b)(6)).(h3) the revision reported was likely the result of osteolysis which may have been due to patient conditions and/or the reported prosthesis wear.The wear was likely due to being implanted for approximately 15.5 years.However, this cannot be confirmed because the component was not returned for evaluation.The most probable root cause associated with the reported event of "osteolysis¿ is associated with destruction or disappearance of bone tissue likely related to weakened integration of an implant at the bone-implant interface.Furthermore, the most probable root cause associated with the reported event of "prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.Section d1: brand name section d4: model number, catalog number, serial number, expiration date, unique identifier (udi) # section g4: pma/510(k)number.Section h4: device manufacture date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUMATCH GXL 15DEG LINER 32MM SZ H
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key13578425
Report Number1038671-2022-00212
Device Sequence Number5539710
Product Code JDI
UDI-Device Identifier10885862023254
UDI-Public10885862023254
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K051556
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2010
Device Model NumberACUMATCH GXL 15DEG LINER 32MM SZ H
Device Catalogue Number132-32-28
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer06/29/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/02/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-