| Model Number |
110-210 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/25/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: analysis of the returned device identified: opening curved on posterior, underweight and brown encapsulation total in the device.A visual and microscopic analysis was performed which identified: sharp opening on posterior assessed as a surgical impact opening, for the stress mark in the shell and wear abrasion.A dimension measurement in the shell was performed which identify the thickness within specification based on the device analysis the final assessment is: a sharp opening on posterior assessed as a surgical impact opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device has been explanted.
|
| |
|