Model Number AX1000
Medical Device Problem Code Application Program Problem: Dose Calculation Error (1189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The transfer set involved has been received for evaluation.Device evaluation not yet completed.The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: it was observed the transfer set had what appears to be residue in it from the lipids used that day in production.No injuries were reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The transfer set involved has been received for evaluation.Device evaluation not yet completed.The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: it was observed the transfer set had what appears to be residue in it from the lipids used that day in production.No injuries were reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical, inc.Internal report (b)(4).The actual device involved in the reported incident was not returned for evaluation.R&d investigation show that the logs are incomplete.Both folders in b.Docs have the same log files, which only go until 11:40am on jan24th, and do not show any compounded orders or even any dispenses at all.The issue was reported that "at the end of production that day when the final patient order had completed, the staff noticed what appeared to be residue left over from the lipids hung that day" since there are no compounded orders, there is nothing close to the time that they were concerned about residual lipids.All that can be seen is that they pass leak check on the first attempt, and what their template initially was (though they might have modified it by the time the issue occurred) with the knowledge that the sequence might have changed, and without knowing the details of the order that may have caused issues, some notes on the default template sequence: 1) there is no indication that the template is not ideal for lipid residue.Lipid is not pumping last, swfi comes after it.2) if there was concern about the visible residue being ca/phos interaction instead of lipid, they have a ui buffer on kphos and naphos to begin the day.This means that there should be adequate separation of ca/phos to minimize any possible interaction.Again, these buffers could have been removed later in the day.At this point (4/11/2022) , if the logs haven't been gathered or retained by vanderbilt, the system would have cleared them because the panel only stores the last 60 days of software logs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APEX¿
Common Device Name
SYSTEM/DEVICE, PHARMACY CO
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
1601 wallace drive, suite 150
carrollton TX 75006 6690
Manufacturer (Section G)
B. BRAUN MEDICAL INC
1601 wallace drive, suite 150
carrollton TX 75006 6690
Manufacturer Contact
jonathan severino
861 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key13578760
Report Number2523676-2022-00030
Device Sequence Number4368999
Product Code NEP
UDI-Device Identifier04046955048502
UDI-Public(01)04046955048502
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAX1000
Device Catalogue NumberAX1000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2022
Supplement Date Received by Manufacturer01/26/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date10/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-