| Model Number |
1999/40 |
| Medical Device Problem Codes |
Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented in clinic with twiddlers syndrome.The right atrial (ra) lead was found to be dislodged.Repositioning of the ra lead was attempted, but a stylet failed to be advanced into the ra lead.The ra lead was explanted and replaced.Patient condition was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported events were dislodgement, twiddlers syndrome, and stylet insertion test.As received, a complete lead was returned for analysis.Visual examination of the lead found the helix retracted and clogged with blood / tissue.After cleaning and applying torque directly to the connector pin, the helix could be extended and retracted.The full helix extension length was measured within specification.Analysis was normal.No anomalies were found.
|
| |
|