| Model Number |
FX-410615 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/10/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: opening, creases fold, wear abrasion and weight to the spec,.A microscopic analysis was performed which identify a crease sharp opening on radius.Based on the device analysis the final assessment is: a crease sharp opening on radius assessed as to a fold flaw opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side rupture.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|