Event or Problem Description
On (b)(6) 2021 after monitoring had ended, an unexpected event occurred during removal of the probe.The crystal did not detach as normal.The wire broke and remained in the patient's body.On (b)(6) 2022 a surgeon performed an additional operation to remove the crystal probe.The patient was discharged from the hospital normally and smoothly.
 
Additional Manufacturer Narrative
Product code: itx.Pma/510(k): k171272.The event is currently under investigation.A follow up report will be submitted upon receipt of additional information or completion of the investigation.
 
Additional Manufacturer Narrative
D2b: product code: itx.G5: pma/510(k): k171272.The device was not returned on this complaint, therefore a physical investigation could not be performed and the customer's complaint could not be confirmed other than by their testimony.Per complaint: "on (b)(6) 2021, after the monitoring ended, an unexpected occurred when the probe wire was removed (the crystal sensor did not detach from the silicone cuff as normal, but the wire was interrupted, and part of the wire remained in the patient's body), we had a suggestion at that time the crystal probe should be removed.The surgeon also performed another operation to remove the crystal probe on (b)(6) 2022, and the patient was discharged from the hospital normally and smoothly." the device history record (dhr) was reviewed, including manufacturing and quality control records and there are no signs to indicate that the device was not manufactured to current specifications.This complaint mode is being monitored, tracked and trended per the cvi post market surveillance and complaint handling processes.A risk assessment will be performed and documented in the complaint summary tab in trackwise.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK-SWARTZ DOPPLER PROBE
Manufacturer (Section D)
COOK VANDERGRIFT INC
1186 montgomery lane
vandergrift PA 15690
Manufacturer (Section G)
COOK VANDERGRIFT INC
1186 montgomery lane
vandergrift PA 15690
Manufacturer Contact
brian johnston
1186 montgomery lane
vandergrift, PA 15690
7248458621
MDR Report Key13580118
Report Number2522007-2022-00005
Device Sequence Number5526771
Product Code ITX
UDI-Device Identifier00827002213630
UDI-Public(01)00827002213630(17)240630(10)N181192
Combination Product (Y/N)N
Initial Reporter CountryTW
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Model NumberG21363
Device Catalogue NumberDP-SDP001
Device Lot NumberN181192
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer01/25/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-