• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE TABLET; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE TABLET; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10027
Medical Device Problem Code Erratic or Intermittent Display (1182)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a cori assisted tka surgery, the real intelligence tablet and the real intelligence 24 in.Touch screen went black every two seconds, no matter were they put them and no matter at what stage of the surgery they were.After a few times, while waiting for the screens to work, the real intelligence cori froze and they had to restart and then they received a system critical error.They used the main monitor to complete the procedure with a significant delay.The patient was not harmed beyond the reported problem.
 
Additional Manufacturer Narrative
The cori tablet used in treatment, was returned for evaluation.A relationship between the reported event and the device was established functionality.As soon the tablet was connected to the cori console it started malfunctioning.During evaluation, the tablet never display content.It will power on (green light), in less than 30 seconds went go to sleep (orange light), approximately 30 second later turned off and all the connection indicators in the console turned off for approximately 30 seconds.This failure was repetitive all the time the tablet was connected to the console.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of this case.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.The most likely cause of this event is an electrical failure that affect the power cycle of the tablet.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored through complaint investigation and post market surveillance activities.Internal complaint reference number: case-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE TABLET
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13580409
Report Number3010266064-2022-00121
Device Sequence Number4369024
Product Code OLO
UDI-Device Identifier00885556757451
UDI-Public00885556757451
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberROB10027
Device Catalogue NumberROB10027
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2022
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date03/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1