| Model Number |
890919 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
01/23/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The customer reported the lack of flexibility of the splint which caused a perforation of the nasal septum with necrosis of the cartilage.No additional information has been provided to date.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device has not been returned to olympus for evaluation.The investigation is in process.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the manufacturing date, h4 updated.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer¿s investigation and the results of the device history records (dhr) review.New information added to h6, h10.The dhr for this device were reviewed and all records indicated the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is likely due customer use.No causal relationship was identified between the event and device.
|
| |
|
Search Alerts/Recalls
|