| Catalog Number |
321-07-42 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during impaction, the resident impacted the humeral liner trial into the humeral adapter tray to get it seated, utilizing a 42mm humeral liner impactor tip.Upon impaction, the top of the impactor tip fracted off.There was no surgical delay.Patient was last known to be in stable condition following the event.Device to be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during impaction, the resident impacted the humeral liner trial into the humeral adapter tray to get it seated, utilizing a 42mm humeral liner impactor tip.Upon impaction, the top of the impactor tip fracted off.There was no surgical delay.Patient was last known to be in stable condition following the event.Device to be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) based on historical complaint investigations and the returned device, the fractured humeral liner impactor tip reported was likely the result of a stress riser introduced during impaction, which led to crack initiation, propagation, and ultimate fracture.Additionally, the devices may have experienced material degradation if sterilized beyond the recommended parameters listed in the reprocessing instruction which could have led to the observed failure.
|
| |
|