| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that during an emergency care patient procedure, using a glidescope bflex 3.8 single-use bronchoscope, the light on the bronchoscope went out.The customer attempted to unplug and re-plug the bronchoscope, but the light would not turn back on.The customer completed the procedure using a different manufacturer device.No harm to the patient or user was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's glidescope bflex 3.8 single-use bronchoscope was returned to verathon for evaluation.A verathon technical service representative evaluated the returned device and was able to confirm the reported light issue.When connecting the glidescope bflex 3.8 to both a and b ports of a known, good, test monitor the led failed to illuminate.The led failure was also observed when the quick-connect cable ends were reversed.Upon completion of the evaluation, the subject glidescope bfflex 3.8 was scrapped and the customer was provided a replacement device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
| |
|