ARROW INTERNATIONAL LLC ARROW CVC SET: 3-LUMEN 8.5FR X 20CM; CATHETER, PERCUTANEOUS Back to Search Results
Model Number IPN030779
Medical Device Problem Codes Deformation Due to Compressive Stress (2889); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported it was difficult to advance the catheter on the guide.The catheter was blocked and that kinked the guide.The catheter was removed and the same guide was used to insert a new catheter at the same insertion site.No patient harm reported.The patient's condition is reported as fine.
 
Event or Problem Description
It was reported it was difficult to advance the catheter on the guide.The catheter was blocked and that kinked the guide.The catheter was removed and the same guide was used to insert a new catheter at the same insertion site.No patient harm reported.The patient's condition is reported as fine.
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed with one relevant finding identified.For material mcz-15853-008a, lot 72c20a0140, a non-conformance was initiated for to address the issue of "swg does not go freely throughout juncture hub".Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 3-LUMEN 8.5FR X 20CM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key13581452
Report Number3006425876-2022-00167
Device Sequence Number4369065
Product Code DQY
UDI-Device Identifier10801902029017
UDI-Public10801902029017
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date11/30/2024
Device Model NumberIPN030779
Device Catalogue NumberCS-15853
Device Lot Number71F20B0830
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date02/22/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-