| Model Number |
1804300-15 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely the device interacted with the heavily calcified, moderately tortuous and 85% stenosed artery causing the reported failure to advance.Continued interaction during retraction of the device caused the reported difficulty to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a heavily calcified, moderately tortuous left anterior descending artery that is 85% stenosed.The 3.0x15mm xience skypoint failed to cross due to anatomy.It was noted resistance during removal with anatomy was felt.A non-abbott device was used to successfully complete the procedure.There was no adverse patient effects and no clinically significant delay.No additional information was provided.
|
| |
|