• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KEYSTONE DENTAL INC RESTO; RESTORE WD 6.0X8 RBM THREADED IMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KEYSTONE DENTAL INC RESTO; RESTORE WD 6.0X8 RBM THREADED IMP Back to Search Results
Model Number RS6008K
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/25/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Failure to osseointegrate.
 
Additional Manufacturer Narrative
Keystone dental inc.Provides labeling with all implants.Failure to osseointegrate and loss of osseointegration (implant mobility) is identified as an adverse reaction and identified in the list of warnings in all endosseous dental implant labeling.The specific cause for a particular complaint is often not readily identified as there are various factors that contribute to the risk of implant failure.These include patient factors such as prior oral infection, poor bone quality or quantity, systemic conditions such as diabetes and uncontrolled hypertension.Technique factors such as overheating of the implant site may cause necrosis of the bone, preventing growth (pelayo et al., 2008; turkyilmaz & al., 2008).Patient habits such as tobacco use (heitz-mayfield & huynh-ba, 2009), alcohol or drug abuse, poor oral hygiene, and bruxism (salvi & braegger, 2009) may also lead to implant failure.In addition, improper surgical technique can lead to implant failure and/or loss of supporting bone (salvi & braegger, 2009).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTO
Common Device Name
RESTORE WD 6.0X8 RBM THREADED IMP
Manufacturer (Section D)
KEYSTONE DENTAL INC
154 middlesex turnpike
burlington MA 01803
Manufacturer (Section G)
KEYSTONE DENTAL INC
154 middlesex turnpike
burlington MA 01803
Manufacturer Contact
ryan raymond
154 middlesex turnpike
burlington, MA 01803
7813283676
MDR Report Key13581949
Report Number3005990499-2022-10585
Device Sequence Number4555995
Product Code DZE
UDI-Device IdentifierD768RS6008K0
UDI-Public+D768RS6008K0/$$323113042695K
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 02/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/29/2023
Device Model NumberRS6008K
Device Catalogue NumberRS6008K
Device Lot Number42695
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Initial Report FDA Received Date02/22/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/17/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age58 YR
Patient SexMale
-
-