| Model Number |
PCE45A |
| Medical Device Problem Codes |
Failure to Fire (2610); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.Additional information was received: this thorax was an emergency.The stapler did the first normal stapling and then locked the mandible.The surgeon and his team had no way to open it and it was necessary to change it.Initial staple line was normal and surgery went normally with stapler replacement.There is no further news of the patient's condition.Additional information has been requested: did the device deliver any staples? if yes, were the staples formed properly? if yes, was the staple line complete? did the device cut? if yes, was the cut line complete? if yes, was there any issue with the cut line such as jagged, dull knife, irregular, etc? was the device locked on tissue with the jaws closed? if yes, how was the device removed? what troubleshooting steps were attempted to open the device (e.G.Squeezing the closing trigger while simultaneously depressing the anvil release button, knife reverse switch, manual override)? did the jaws eventually open? is the current patient status known? was there any patient consequence or change in the post-operative care of the patient as a result of the event? (extended hospital stay, readmission, re-operation, etc.) attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.An analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances / manufacturing irregularities were identified.As part of ethicon's quality process, all devices are manufactured, inspected, and distributed to approved specifications.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure, videolaparoscopy segmentectomy, the surgeon identified surgical stapler failure.The same reported that it was not firing, then locked and was unable to undock.A new stapler has been requested to continue the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 5/4/2022.D4: batch # t5ey5l.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample determined that one pce45a reload was received with no apparent damaged and with an ecr45m reload loaded.The reload was received partially fired 1/16.The manual override door out of position; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.The bailout system was reset and then, the returned device and reload were tested for functionality in the straight position by resetting and reloading it into the device.The device achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples meet the staple form release criteria.It is possible that while loading the reload, the reload was pushed farther back than the reload alignment stop windows resulting in the knife pushing the one piece sled forward and locking the reload.Please reference the instruction for use for more information.The device opened and closed without any difficulties noted.The instructions for use do contain the following caution: if the instrument is partially fired, remove the instrument and slide the knife reverse switch forward to return the knife to home position.To open the jaws, squeeze the closing trigger, then simultaneously press the anvil release button on either side of the instrument.While pressure is still on the anvil release button, slowly release the closing trigger.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|