MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE FIXED WIRE; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE FIXED WIRE; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558340
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2021.During the procedure, it was noted that the balloon popped while being inflated to 9mm.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: device code a0402 captures the reportable event of balloon burst.Block h10: investigation results: a visual examination of the returned complaint device found that the balloon was torn longitudinally, which does not confirm the reported event of balloon popped; however, the balloon torn problem found could have been interpreted by the customer as the reported event of balloon burst.Microscopic analysis was performed, no damage found on the catheter of the device.Based on the available information, it is possible that factors encountered during the procedure, such as interaction with any surface during the procedure could create friction on the balloon, causing the problem of torn longitudinal during the procedure.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was used per the instructions for use (ifu)/product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on november 29, 2021.During the procedure, it was noted that the balloon popped while being inflated to 9mm.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2021.During the procedure, it was noted that the balloon popped while being inflated to 9mm.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h2: additional information: blocks d7 (sud reprocessed and reused?), block e1 (initial reporter title, initial reporter first name and last name ), and h8 (usage of device) block h6: device code (b)(6) captures the reportable event of balloon burst.Block h10: the device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE FIXED WIRE
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13583420
Report Number3005099803-2022-00895
Device Sequence Number20244983
Product Code KNQ
UDI-Device Identifier08714729201991
UDI-Public08714729201991
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K971320
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/21/2022
Device Model NumberM00558340
Device Catalogue Number5834
Device Lot Number0026048431
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/09/2022
Supplement Date Received by Manufacturer02/23/2022
04/04/2022
Initial Report FDA Received Date02/23/2022
Supplement Report FDA Received Date03/16/2022
04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-