• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. INFUSOMAT®; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. INFUSOMAT®; SET, ADMINISTRATION, INTRA Back to Search Results
Model Number 480258
Medical Device Problem Code Device Misassembled During Manufacturing /Shipping (2912)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: brief inquiry description: green slide clamp on wrong side of pump segment.Detailed inquiry description: while priming, the nurse noticed that the green clamp was on the wrong side of the pump segment.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: brief inquiry description: green slide clamp on wrong side of pump segment.Detailed inquiry description: while priming, the nurse noticed that the green clamp was on the wrong side of the pump segment.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4) and it is being submitted as a correction.*correction, the investigation findings code was submitted incorrectly.The correct investigation findings code is 170 (manufacturing process problem identified).
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) sample without packaging was returned for evaluation.The sample was visually evaluated; it was confirmed that the horizon cassette was assembled incorrectly.The reported defect was confirmed.A one-year historical review was performed and it should be noted that this defect is the only reported case.Incidents of this nature are attributed to operator oversight during the assembly of the product.The outlook pump cassette is manually assembled onto the tubing.The operator could have overlooked the position in which they assembled the cassette.Although our training procedures ensure that all our employees are properly trained in their areas of responsibility, an oversight on the part of the operator can attribute to an incident of this nature.As a result of this occurrence a quality notification and retrain was generated for all associates involved in the hand assembly process.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOMAT®
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer (Section G)
B. BRAUN DOMINICIAN REPULIC INC.
las americas industrial park
km22 autopista las americas
santo domingo
DR  
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key13583869
Report Number2523676-2022-00089
Device Sequence Number5731220
Product Code FPA
UDI-Device Identifier04046955324835
UDI-Public(01)04046955324835
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K142036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Model Number480258
Device Catalogue Number480258
Device Lot Number0061791069
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer02/04/2022
02/04/2022
Initial Report FDA Received Date02/23/2022
Supplement Report FDA Received Date04/16/2022
04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-