• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP - SUPERDIMENSION INC EDGE; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP - SUPERDIMENSION INC EDGE; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number SD300LG
Medical Device Problem Codes Failure to Advance (2524); Application Program Problem (2880)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Out of an abundance of caution, superdimension is filing this mdr due to the additional risk associated with multiple exposures to general anesthesia.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the procedure, the locatable guide (lg) did not extend beyond the extended working channel (ewc).The patient was already under anesthesia when the procedure was cancelled.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the failure could not be determined.It was reported that the patient procedure was cancelled or aborted due to a system issue, and the lg length is too short resulting in inadequate protrusion out from the extended working channel.The reported issues could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGE
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer (Section G)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer Contact
amy beeman
161 cheshire lane, suite 100
plymouth, MN 55441
7632104064
MDR Report Key13583948
Report Number3004962788-2022-00009
Device Sequence Number5539866
Product Code JAK
UDI-Device Identifier10884521199514
UDI-Public10884521199514
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151376
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2024
Device Model NumberSD300LG
Device Catalogue NumberSD300LG
Device Lot Number516890
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date02/23/2022
Supplement Report FDA Received Date04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-