| Model Number |
BOPS4313 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Doctor reported fracture of abutment screw at tooth site 25.Unable to removed screw, implant was also removed.Doctor's assistant confirmed by phone that implant fractured when removing the screw.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Additional device information unknown / not provided.Email address unknown / not provided.Device manufacturer date unknown / not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).One certain® low profile one-piece abutment 3.4mm(d) x 2mm(h), ilpc342u & one 3i t3® parallel walled implant 4/3 x 13mm, bops4313 was returned for investigation.Visual inspection via naked eye and camera magnification was completed.The abutment was fractured at the screw threads.Damage observed on the returned implant due to removal is not considered a malfunction or failure of the device.Zimvie is not responsible for any damage caused by the clinician¿s removal of the device.Device history record (dhr) review was completed for the subject lot number (2016100964).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (2012030146) for similar event (complaint category keywords: fracture screw) and revealed no existing non-conformances/capa/hhe/d/ie/product holds for the reported product.Based on the available information, device malfunction did occur and the reported event was confirmed following inspection and evaluation.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of the report.
|
| |
|
Search Alerts/Recalls
|