Initial reporter is synthes sales representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, the patient underwent a procedure for fracture of distal radius.During the surgery, the sleeve did not fit into the hole located at the distal center of the guiding blocks.The surgery was completed successfully with a thirty (30) minute delay.No further information is available.This report is for one (1) guide block for 2 column plate 7 hole head/right.This is report 1 of 3 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: a product investigation was conducted.The product was returned to us cq for evaluation.The us cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the guidingblock f/2-column dist-rad-pl2.4 7 had three of the holes deformed openings.Nicks and scratches were observed on the device indicating signs of use.No other issues were identified with the returned device.No new additional information can be added after reviewing the provided images.The dimensional inspection was performed and the concerned holes were found to be non-conforming due to post manufacturing damage.A functional test was performed and it was confirmed that the drill sleeve could not pass smoothly through the deformed holes of the guiding block.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the guidingblock f/2-column dist-rad-pl2.4 7 was observed to have deformed holes.While no definitive root cause could be determined for the reported complaint condition from the available information, it is probable that the holes of the guidingblock f/2-column dist-rad-pl2.4 7 were deformed due to the exposure to excessive/unintended forces.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h6: a device history record (dhr) review was conducted: part: 03.111.700; lot: 14-8841; manufacturing site: selzach; supplier: (b)(4); release to warehouse date: 01 sep 2014; expiration date: n/a.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.E3: reporter is a healthcare professional.
|