Date of event is an unknown date in 2022.This report is for an unknown screw/unknown lot.Part and lot number are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that on (b)(6) 2022, patient underwent revision surgery due to two broken expedium cortical screws at the s1 level.Patient and procedure outcome were unknown.Patient had a previous l3-s1 fusion, which was extended up to l2 and added sai after replacement of the broken screws.Procedure outcome and patient status are unknown.This report is for an unknown expedium screw.This is report 1 of 2 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: h3, h6: a photo investigation was completed: visual analysis of the provided photos revealed that both screws are broken near the tip.The photos quality was poor, therefore no further details could be noted.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.Given that no lot number was provided, a manufacturing record evaluation (mre) review cannot be performed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the screws.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|