ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number MF-410640
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Seroma (2069); Lymphoma (3263); Breast Implant Associated Anaplastic Large Cell Lymphoma (BIA ALCL) (4549)
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported unknown side "confirmed alcl." healthcare professional additionally reported right side rupture, and seroma.Pathological marker cd30+ has been received.Pathological marker alk- has not been received.The device remains implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The events of lymphoma - alcl and seroma-late are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture, seroma-late, and lymphoma-alcl.
 
Event or Problem Description
Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: analysis of the returned device identified the weight the device within specification, black particles in the shell and deformation.Cloudy and voids in the gel observed after autoclave cycle.Based on the device analysis the final assessment is: no issues found related with the manufacturing.
 
Event or Problem Description
Healthcare professional reported right side "confirmed alcl", "rupture", "seroma".Pathological marker cd30+ has been received.Pathological marker alk- has not been received.The device has been explanted.
 
Event or Problem Description
Healthcare professional reported unknown side "confirmed alcl." healthcare professional additionally reported right side rupture, and seroma.Pathological marker cd30+ has been received.Pathological marker alk- has not been received.The device has been explanted.
 
Additional Manufacturer Narrative
Previous medwatch submission noted lymphoma-alcl.Upon review of received pathology reports, allergan medical safety determined that it should be reported as lymphoma-alcl-suspected.Only pathology marker cd30+ was provided.Alk- was not received.Upon review of additional information, the recall number previously reported in h.7./h.9.Should not have been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key13585062
Report Number9617229-2022-02354
Device Sequence Number5539917
Product Code FTR
UDI-Device Identifier10888628003521
UDI-Public10888628003521
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/07/2020
Device Model NumberMF-410640
Device Catalogue NumberMF-410640
Device Lot Number2773964
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer04/12/2022
05/25/2022
01/12/2024
Initial Report FDA Received Date02/23/2022
Supplement Report FDA Received Date05/11/2022
06/22/2022
02/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2011068-7/2/19-001-R
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-