Model Number 1006
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/03/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that tip detachment occurred.The target lesion was located in the mildly tortuous and severely calcified pedal artery (pa).A savion flx guidewire was advanced to gain access from the foot up to the peroneal artery.However, during the procedure, a bit of resistance was felt upon attempting to advance the device and the entire floppy tip of the guidewire broke off in pa and was left behind.The wire was removed but the tip of the guidewire had to be left in the patient's foot due to the size of the vessel and no device or snare small enough to attempt a retrieval.The procedure was completed with another savion guidewire.No patient complications were reported.
 
Event or Problem Description
It was reported that tip detachment occurred.The target lesion was located in the mildly tortuous and severely calcified pedal artery (pa).A savion flx guidewire was advanced to gain access from the foot up to the peroneal artery.However, during the procedure, a bit of resistance was felt upon attempting to advance the device and the entire floppy tip of the guidewire broke off in pa and was left behind.The wire was removed but the tip of the guidewire had to be left in the patient's foot due to the size of the vessel and no device or snare small enough to attempt a retrieval.The procedure was completed with another savion guidewire.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: unit returned, overall visual revision did not identify failures or evidence that could be lost due to decontamination process.The device was inspected according to procedure: the returned guidewire had the distal tip portion with polymer section detached.No other issues were found.The device was measured in three sections (distal - medium - proximal).The distal, medium and proximal outer diameter were within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAVION FLX
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13586585
Report Number2134265-2022-01373
Device Sequence Number12173268
Product Code DQX
UDI-Device Identifier08714729978817
UDI-Public08714729978817
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K180726
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/03/2023
Device Model Number1006
Device Catalogue Number1006
Device Lot Number0027964289
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer03/03/2022
Initial Report FDA Received Date02/23/2022
Supplement Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-