• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION COMPONENT; ANESTHESIA CONDUCTION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION COMPONENT; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Model Number IPN040153
Medical Device Problem Codes Material Fragmentation (1261); Material Too Soft/Flexible (4007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It seems that the catheter is very elastic until you reach the markings then it is brittle and breaks when pulled.It can be repeated with other catheters that were not used on a patient.I believe two patients, yes standard epidurals for or surgery, no injury thus far.
 
Additional Manufacturer Narrative
Qn# (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Additional Manufacturer Narrative
(b)(4).A device history record review could not be performed as no lot number was provided by the customer.A review of sales history data was performed to obtain a lot number.However, there was no record of purchase of this product by this customer.A review of design change history for part number kz-05400-030 was performed as a part of this investigation.No design changes have been made to this part in the past two years that would have led to this complaint.A corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.The device history records were not reviewed as no lot number was provided by the customer and no sales history records could be found for this customer for this product.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without a sample.No further action is required at this time.
 
Event or Problem Description
It seems that the catheter is very elastic until you reach the markings then it is brittle and breaks when pulled.It can be repeated with other catheters that were not used on a patient.I believe two patients, yes standard epidurals for or surgery, no injury thus far.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION COMPONENT
Common Device Name
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key13586785
Report Number3006425876-2022-00201
Device Sequence Number20001951
Product Code BSO
UDI-Device Identifier10801902139679
UDI-Public10801902139679
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K140110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN040153
Device Catalogue NumberEC-05500
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer03/31/2022
Initial Report FDA Received Date02/23/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-