| Catalog Number |
03.010.101 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: the complaint device (drill bit ø4.2 calibr l145 3flute w/coup) was not received.The image was reviewed, and the complaint condition was confirmed.After review of the photo provided, it was noted that the drill tip was broken.A manufacturing record evaluation was performed for the finished device [u358119] number, and no non-conformances were identified.A definite assignable root cause could not be determined based on the provided information.As the device was not returned, and as-received condition could not be assessed and a dimensional inspection and document/specification review were not completed.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: part #: 03.010.101, lot #: u358119, manufacturing site: (b)(4), supplier: (b)(4), release to warehouse date: 04 may 2020, expiration date: n/a.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, during an unknown procedure, when performing the bone drilling for the placement of the distal pin nail, the drill bit fractured in its proximal part.No further information was provided.This report is for one (1) 4.2mm three-fluted radiolucent drill bit/needle point/145mm.This is report 1 of 1 for (b)(4).
|
| |
|
Event or Problem Description
|
|
There was no surgical delay.A fragment remains in the patient, no other surgical procedure is required for removal.The procedure was successfully completed.There was a satisfactory patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|