• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS Back to Search Results
Model Number AR40E
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Weight and ethnicity: unknown/ not provided.If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.(b)(6).The intraocular lens (iol) was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the lagging haptic was missing and found inside the injector during the implantation of the intraocular lens (iol) into the patient's operative eye.Lens was fully implanted.However, as the haptic was broken, the iol was removed during the same procedure.There were incision enlargement and sutures performed.There was a delay in procedure of 20 minutes.No other intervention was reported.No further information was provided.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes.Date returned to manufacturer: mar 16, 2022.Section h3.Device evaluated manufacturer? yes.Device evaluation: visual inspection under magnification revealed that the lens was received with a haptic detached.The lens was cleaned, and no further issues were observed.Conclusion: as a result of the investigation there is no indication of a product quality deficiency or product malfunction and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Corrected data: upon further review it was noted that report type for this report was for product malfunction only instead of for both adverse event and malfunction as noted in the initial mdr submitted.Also, it was inadvertently missed to report the pre-operative visual acuity which was actually counting fingers.This report is submitted as correction.Additional information: new information received that as patient left aphakic, it was confirmed that the patient was implanted with another iol sometime later.The iol was stable and in place.The vision of patient was stable.The patient post-operative visual acuity was 20/100, (patient is also s/p repair of chronic retinal detachment).No further information was provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13588541
Report Number3012236936-2022-00461
Device Sequence Number20288552
Product Code HQL
UDI-Device Identifier05050474501973
UDI-Public(01)05050474501973(17)260809
Combination Product (Y/N)N
Initial Reporter CountryRP
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/09/2026
Device Model NumberAR40E
Device Catalogue NumberAR40E00135
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2022
Supplement Date Received by Manufacturer04/08/2022
05/03/2022
Initial Report FDA Received Date02/23/2022
Supplement Report FDA Received Date04/25/2022
05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
-
-