• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS EN SNARE; PERCUTANEOUS VASCULAR SNARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS EN SNARE; PERCUTANEOUS VASCULAR SNARE Back to Search Results
Catalog Number EN2007030/B
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The account alleges that during an inferior vena cava filter removal procedure on a referral patient, the tip of a vascular snare device detached within the patients ivc.A second vascular snare device was prepped and deployed in attempts to remove the primary foreign body from the patient when the tip of the second snare device also detached.The physician states that usual resistance was experienced during the removal procedure even though the filter was embedded within the wall of the ivc.Original implant date in unknown.No additional consequences to report.
 
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and one similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EN SNARE
Common Device Name
PERCUTANEOUS VASCULAR SNARE
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS
parkmore business park west
ballybrit, galway
EI 
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS
parkmore business park west
ballybrit, galway
EI  
Manufacturer Contact
bryson heaton bsn,rn.
1600 merit parkway
8012084662
MDR Report Key13588588
Report Number9616662-2022-00005
Device Sequence Number14940029
Product Code MMX
UDI-Device Identifier00884450452479
UDI-Public884450452479
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K151497
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/09/2024
Device Catalogue NumberEN2007030/B
Device Lot NumberK2254244
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer04/07/2022
Initial Report FDA Received Date02/23/2022
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-