| Catalog Number |
EN2007030/B |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The account alleges that during an inferior vena cava filter removal procedure on a referral patient, the tip of a vascular snare device detached within the patients ivc.A second vascular snare device was prepped and deployed in attempts to remove the primary foreign body from the patient when the tip of the second snare device also detached.The physician states that usual resistance was experienced during the removal procedure even though the filter was embedded within the wall of the ivc.Original implant date in unknown.No additional consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and one similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|