Catalog Number NCEUP3512X |
Device Problems
Deflation Problem (1149); Difficult to Remove (1528)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 02/12/2022 |
Event Type
Injury
|
Event Description
|
An attempt was made to use a nc euphora rx balloon catheter to treat lesion with 99% stenosis in the distal right coronary artery (rca).The device was inspected with no issues.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device passed through a previously-deployed stent.Excessive force was not used during delivery.It was reported that balloon deflation difficulties were encountered, the device would not deflate at the lesion site.The balloon could not be retracted and taken out.An attempt was made to pull the device back to puncture the device with the guide wire, but was unsuccessful.The patient was transferred to a higher-level hospital for thoracotomy treatment and the device was removed.No further patient injury was reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: the lesion was severely calcified.There were no issues noted when removing the protective sheath and the packaging stylette.Resistance was not encountered when advancing the device.No difficulties were found during device inflation.Deflation difficulties occurred after the first inflation when the balloon was inflated to 12 atm for 6 seconds.The patient is still in the icu.Correction: initial reporter address and phone number.Annex a code added medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|