| Model Number |
C111 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable ise indirect gen.2 sodium results from the cobas c 111 with ise analyzer.The samples were repeated on a cobas c501 analyzer at another site.Patient 1 initial result was 132 mmol/l and the repeat result was 140 mmol/l.Patient 2 initial result was 129 mmol/l and the repeat result was 137 mmol/l.Patient 3 initial result was 132 mmol/l and the repeat result was 140 mmol/l.It was requested but was not known if any of the questionable results were reported outside of the laboratory.The electrode lot number and expiration date were requested but were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|