ENCORE MEDICAL L.P. ALTIVATE REVERSE SHOULDER; AR, SMALL SOCKET INSERT, 40MM SEMI CONSTRAINED EPLUS Back to Search Results
Model Number 509-03-040
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to instability.
 
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as instability.The previous surgery and the surgery detailed in this event occurred 1 year and 3 months apart.The healthcare professional indicated there was a significant adverse event to the patient and there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to instability.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE REVERSE SHOULDER
Common Device Name
AR, SMALL SOCKET INSERT, 40MM SEMI CONSTRAINED EPLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
kiersten soderman
9800 metric blvd
austin, TX 78758-5445
MDR Report Key13590877
Report Number1644408-2022-00259
Device Sequence Number5912581
Product Code KWS
UDI-Device Identifier00190446257646
UDI-Public(01)00190446257646
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K172351
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/06/2023
Device Model Number509-03-040
Device Catalogue Number509-03-040
Device Lot Number960W1021
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2022
Supplement Date Received by Manufacturer02/28/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date03/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/13/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexMale
-
-