| Catalog Number |
11447475160 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable glucose results from accutrend plus meter serial number (b)(4).At 8:20am the result was 108 mg/dl.At 8:25 am the result was 124 mg/dl.At 8:34 am the result was 141 mg/dl.The same fingerstick was used to obtain all three tests.None of the results were believed to be correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The strips were requested for investigation.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
No material was returned by the customer.Retention samples of glucose lot 48192609 were measured with retention controls and 3 edta blood on reference devices which were compared to a laboratory method.All results were within the control range.No significant differences from reference devices to lab method.
|
| |
|