SMITH & NEPHEW, INC. T25 SELF-RET SCREWDRIVER W/QC 178MM Back to Search Results
Model Number 71173616
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during total hip replacement, the tip of a t25 self-ret screwdriver w/qc 178 mm broke off inside the patient.All pieces were recovered.Surgery was resumed, without any delay, with a smith & nephew back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during total hip replacement, the tip of a t25 self-ret screwdriver w/qc 178 mm broke off inside the patient.All pieces were recovered.Surgery was resumed, without any delay, with a smith & nephew back-up device.Patient was not injured as consequence of this problem.
 
Event or Problem Description
It was reported that, during total hip replacement, the tip of a t25 self-ret screwdriver w/qc 178mm broke off inside the patient.All pieces were recovered by hand.Surgery was resumed, without any delay, with a smith & nephew back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be fractured along the tip, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.The clinical/medical evaluation concluded that based on the information provided, the broken tip of a t25 self ret screwdriver w/qc 178mm was removed from inside of the patient.The two undated, unlabeled photos of the device provided supports the complaint.According to the report, the surgeon completed the procedure without delay, using a smith & nephew back up device.Since there was no alleged harm to this patient, no further clinical/medical assessment is warranted at this time.Should any additional relevant patient information be provided, this case would be re-assessed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.B5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T25 SELF-RET SCREWDRIVER W/QC 178MM
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13591603
Report Number1020279-2022-00850
Device Sequence Number11791954
Product Code HXX
UDI-Device Identifier03596010637178
UDI-Public03596010637178
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71173616
Device Catalogue Number71173616
Device Lot Number13KBG0009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer04/05/2022
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/15/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-